HomeProductsInstrumentsU Serials4in,PDM
4in,PDM

U 920G

  • 4in,PDM
  • U 920G is a PDM analog source specifically designed for electroacoustic testing. With its excellent electrical characteristics and superior cost-effectiveness, it is highly suitable for testing digital PDM demodulators and can form a closed loop with the U926 or U920A.

  • Description
  • Download

Description

Specifications


Parameters

Full-load Power Consumption

5V/400mA

Number of USB Shuttle Channels

4

External Analog Channels(Default upon Power-on 

4

PDM Voltage

1.8V

Bus

BNC

Product dimension mm

96*33*140

Working Temperature

-20℃ - 50℃


AO Channel Characteristic Parameters


Idle noise(dBFS)

Sample Rate

fs=48kS/s

Typical Value

-73dBFS
PS:[1]Input signal is 0Vrms or input is short-circuited

       [2]Weighting Method: A-weighted

THD (dB)

Sample Rate

fs=48kS/s

Typical Value

-81dB
PS:[1]Single-ended input, AC-coupled. Input signal is a 1kHz sine wave at 0.5Vrms (digital signal level at -10dBFS)

       [2]Weighting Method: A-weighted

THD+N (dB)

Sample Rate

fs=48kS/s

Typical Value

-78dB
 PS:[1]Single-ended input, AC-coupled. Input signal is a 1kHz sine wave at 0.5Vrms (digital signal level at -10dBFS)

       [2]Weighting Method: A-weighted

Cross talk(dB)

Sample Rate

fs=48kS/s

Typical Value

80dB
PS:[1]Single-ended input, AC-coupled. Input signal is a 1kHz sine wave at 0.5Vrms (digital signal level at -10dBFS)

       [2]Weighting Method: A-weighted


PS:

1.The maximum effective input is 1Vrms (digital signal level at -4dBFS). Inputs higher than this value may result in significant distortion

2.The specifications for 3.5mm input and Shuttle input are the same

3.Calibration values are based on a 1kHz sine wave input with an input amplitude of 0.5VRMS, collected at the digital signal end as -10dBFS


Download