HomeProductsInstrumentsPM Serials4in,DSA
4in,DSA

PM 6180

  • 4in,DSA
  • PM 6180 is a data acquisition card specifically designed for production line testing.

    4- channel analog input ,supporting connections for 2 sets of 3.5mm audio cables to achieve stereo monitoring with adjustable monitoring gain. The card allows enabling or disabling the IEPE constant current source excitation and supports adjustable AC/DC coupling for conditioning microphones, artificial ears, accelerometers, and other IEPE-powered sensors.

    With digital optocoupler isolation, it provides strong anti-interference capability and supports rotational speed signal acquisition.

    It is suitable for acoustic quality analysis, electroacoustic analysis, vibration analysis, NVH analysis, and other demanding measurement applications to capture sound and vibration data.

    Additionally, the PM 6180 offers an open application protocol interface, which is a lightweight and platform-independent interface designed for programming data acquisition from the PM 6180 using your own (or third-party) software.


  • Description
  • Download

Description

Specifications


ParameterValue
模拟采集通道数4
IEPE 通道4路(内置在采集前端)
立体声监听通道4(2组L/R)
IEPE 电源4mA
总线USB-B
产品尺寸 (L*W*H mm) 【不含接口长度】240*270*76.5
接口类型 BNC
适用温度范围-20 - 50℃
工作电压220V
耦合方式AC/DC耦合
Trigger同步触发内部触发延时:1.2ms

Specifications


Parameters

Input Channels

4

IEPE Channels

4 channels (integrated into the front-end acquisition)

IEPE Power

4mA

Stereo Monitoring Channels

4(2 Left/Right)

Trigger Synchronization

Internal Trigger Latency: 1.2 ms

Control

USB-B

Dimension   mm

240*215*76(L*W*H)

Connection Type

BNC

Working Temperature

-20 - 50℃

Coupling methodAC/DC coupling


AI Specifications


AI Parameters

ADC Precision

24 Bit

ADC Type

Δ-Σ

Sample Rate Range    Hz

1k - 100k

FIFO

1024

Mode Of Data Transmission

DMA

Input Range

+/- 10Vrms (Measurement ranges 1Vrms、5Vrms、10Vrms)

Input Impedance

Input Positive Terminal to Ground: 200Kohm

Gain Error

AC coupling+/-0.03 dB       DC coupling +/-0.1 dB   

Flatness20Hz - 20kHz,Typical<+/-0.02 dB Maximum Value 0.03dB     20Hz - 50kHz, Typical<+/-0.02 dB  Maximum Value 0.03dB 

AI Idle noise (uVrms)

Sample Ratefs=48kS/s
fs=96kS/s
Maximum Value15
20
PS:[1]Short-circuit source impedance less than 50 ohms,Operating temperature of 23±5 degrees Celsius,AC coupling,Linear weighting,1Vrms

AI SNR (dB)

Sample Ratefs=48kS/s
fs=96kS/s
Maximum Value115
115

PS:[1]Differential Input,AC Coupling,Input Signal of 1kHz Sine Wave,10Vrms Range,2Vrms

        [2] Bandwidth 22.4k、45k

        [3] Linear Weighting

AI Dynamic Range (dB)

Sample Ratefs=48kS/s
fs=96kS/s
Maximum Value110
109

PS:[1]Differential Input,AC Coupling,Input Signal of 1kHz Sine Wave,5Vrms Range,2Vrms

        [2] Bandwidth 22.4k、45k

        [3] Linear Weighting,SFDR Testing Method

AI THD+N (dB)

Sample Ratefs=48kS/s
fs=96kS/s
Maximum Value-100
-100

PS:[1]Differential Input,AC Coupling,Input Signal of 1kHz Sine Wave,5Vrms Range,2Vrms

        [2] Bandwidth 22.4k、45k

        [3] Linear Weighting

AI Cross talk (dB)

Sample Ratefs=48kS/s
fs=96kS/s
Maximum Value-95-95

PS:[1]Differential Input,AC Coupling,Input Signal of 1kHz Sine Wave,5Vrms Range,2Vrms

        [2] Bandwidth 22.4k、45k

        [3] Linear Weighting


Download